Title: Comparison of Rectal, Axillary, Aural and non-contact thermometers in Febrile children
Authors: Dr Karthick AR, Dr Luke Ravi Chelliah, Dr Theranirajan E, Dr Gangadharan S
DOI: https://dx.doi.org/10.18535/jmscr/v5i6.174
Abstract
Introduction: Highest incidence of febrile illnesses occur in infancy and early childhood and it is essential to determine body temperature accurately, allowing rapid identification of potentially life threatening conditions. Rectal temperature is considered as ‘Gold standard’ but it is inconvenient. This study was designed to assess and compare the performance of three commercially available thermometers, a digital thermometer for axillary and rectal temperature measurement, a non-contact infrared thermometer for forehead temperature measurement and an infrared tympanic membrane thermometer.
Methods and Materials: It was a descriptive study done in febrile children aged 3 months to 5 years, from September 2016 to February 2017 in Department of Paediatrics, Government Medical College and Hospital, Vellore, Tamil Nadu.
Results: Mean rectal temperature was 38.96oC ± 0.62oC. Mean tympanic membrane temperature was 38.80oC ± 0.64oC. Mean axillary temperature was 38.04oC ± 0.62oC. Mean forehead temperature was 38.38oC ± 0.65oC. The co-efficient of correlation between mean rectal and tympanic membrane temperature was 0.958 and between mean rectal and axillary temperature was 0.907. The co-efficient of correlation between mean rectal and forehead temperature was 0.940.
Conclusion: Children experienced more discomfort during rectal thermometry than with others. Amongst the three method, tympanic membrane temperature closely correlated and reflected the rectal temperature.
Keywords: Tympanic membrane temperature; rectal temperature; axillary temperature, non-contact thermometer.
1. Waskewitz S, Berklhamer JE. Outpatient bacteremia: Clinical findings in children under two years with initial temperatures of 39.5o or higher. J Pediatr. 1981;(99):231–3.
2. Terndrup TE, Allegra JR, Kealy JA. A comparison of oral, rectal, and tympanic membrane-derived temperature changes after ingestion of liquids and smoking. Am J Emerg Med. 1989 Mar;7(2):150–4.
3. León C, Rodríguez A, Fernández A, Flores L. Infrared ear thermometry in the critically ill patient. J Crit Care. 2005 Mar 1;20(1):106–10.
4. Sehgal A, Dubey NK, Jyothi MC, Jain S. Comparison of tympanic and rectal temperature in febrile patients. Indian J Pediatr. 2002 Apr 1;69(4):305–8.
5. Bernardo LM, Clemence B, Henker R, Hogue B, Schenkel K, Walters P. A comparison of aural and rectal temperature measurements in children with moderate and severe injuries. J Emerg Nurs. 1996 Oct 1;22(5):403–8.
6. Jean-Mary MB, Dicanzio J, Shaw J, Bernstein HH. Limited accuracy and reliability of infrared axillary and aural thermometers in a pediatric outpatient population. J Pediatr. 2002 Nov 1;141(5):671–6.
7. Chaturvedi D, Vilhekar KY, Chaturvedi P, Bharambe MS. Comparison of axillary temperature with rectal or oral temperature and determination of optimum placement time in children. Indian Pediatr. 2004 Jun;41(6):600–3.
8. Abdulkadir MB, Johnson ’wahab Babatunde Rotimi. A comparative study of rectal tympanic and axillary thermometry in febrile children under 5 years of age in Nigeria. Paediatr Int Child Health. 2013 Aug;33(3):165–9.
9. El-Radhi AS, Patel S. An evaluation of tympanic thermometry in a paediatric emergency department. Emerg Med J EMJ. 2006 Jan;23(1):40–1.
10. Shann F, Mackenzie A. Comparison of rectal, axillary, and forehead temperatures. Arch Pediatr Adolesc Med. 1996 Jan;150(1):74–8.
11. Kistemaker JA, Den Hartog EA, Daanen H a. M. Reliability of an infrared forehead skin thermometer for core temperature measurements. J Med Eng Technol. 2006 Aug;30(4):252–61.
12. Robinson JL, Seal RF, Spady DW, Joffres MR. Comparison of esophageal, rectal, axillary, bladder, tympanic, and pulmonary artery temperatures in children. J Pediatr. 1998 Oct;133(4):553–6.
13. Hooker EA, Smith SW, Miles T, King L. Subjective assessment of fever by parents: comparison with measurement by noncontact tympanic thermometer and calibrated rectal glass mercury thermometer. Ann Emerg Med. 1996 Sep;28(3):313–7.
14. Bliss-Holtz J. Comparison of rectal, axillary, and inguinal temperatures in full-term newborn infants. Nurs Res. 1989 Apr;38(2):85–7.
15. Stewart JV, Webster D. Re-evaluation of the tympanic thermometer in the emergency department. Ann Emerg Med. 1992 Feb;21(2):158–61.
16. Ogren JM. The inaccuracy of axillary temperatures measured with an electronic thermometer. Am J Dis Child 1960. 1990 Jan;144(1):109–11.
17. Wilshaw R, Beckstrand R, Waid D, Schaalje GB. A comparison of the use of tympanic, axillary, and rectal thermometers in infants. J Pediatr Nurs. 1999 Apr;14(2):88–93.
18. Hoffman C, Boyd M, Briere B, Loos F, Norton PJ. Evaluation of three brands of tympanic thermometer. Can J Nurs Res Rev Can Rech En Sci Infirm. 1999 Jun;31(1):117–30.
19. Musumba CO, Griffiths KL, Ross A, Newton CRJC. Comparison of axillary, rectal and tympanic temperature measure-ements in children admitted with malaria. J Trop Pediatr. 2005 Aug;51(4):242–4.
20. Edelu BO, Ojinnaka NC, Ikefuna AN. A comparison of axillary with rectal thermometry in under 5 children. Niger Med J J Niger Med Assoc. 2011;52(4):207–10.