Title: Evaluation of Erythrocyte Sedimentation Rate by an Automated ESR Analyzer Assuming Manual Westergren Method as Gold Standard at S.M.S. Medical College and Hospital, Jaipur (Rajasthan)

Authors: Dr Anand Kumar Verma, Dr Anita Harsh, Dr Kusum Mathur

 DOI: https://dx.doi.org/10.18535/jmscr/v8i7.51

Abstract

Background: The erythrocyte sedimentation rate (ESR) is a widely used simple, inexpensive laboratory test often requested in various clinical conditions. Many automated ESR analyzers are being evolved for measuring ESR to improve efficacy.

Aims and Objective: To evaluate the performance of the Automated ESR Analyzer and to assess and compare accuracy of ESR readings by Automated ESR.

Material and Methods: The 425 samples from 425 consecutive eligible patients fulfilling the inclusion criteria were collected at a tertiary care center in Jaipur and data was analyzed.

Results: Bland and Altman analysis revealed low degree of agreement between two methods especially for ESR values > 25 mm/hr. For all the 425 samples, the mean difference and 95% limits of agreement of both the methods were -0.49±10.09 (95% limits of agreement, -10.58 to 9.60). For ESR values > 25 mm/hr, the 95% limits of agreement was -11.52 to 10.05. For ESR values ≤ 25 mm/hr the 95% limits of agreement was -4.78 to 5.17, which showed a very good agreement between both the methods.

Conclusion: The Automated Analyzer tend to underestimate the Westergren method ESR values which were ≥ 25 mm/hr. The agreement for ESR value ≤ 25 mm/hr was very good. Hence a correction factor should be applied for ESR values while using Automated Analyzer especially for higher ESR values.

Keywords: Comparison, ESR, Westergren, Automated.

References

  1. Plebani M, Piva E. Erythrocyte sedimentation rate: Use of fresh blood for quality control. Am J Clin Pathol. 2002;117:621-6.
  2. Saadeh C. The erythrocyte sedimentation rate: old and new clinical applications. South Med J. 1998;91(3):220-5.
  3. Mahato M, Karkuzhali P. A comparative study of esr by modified westergren method and vesmatic cube 30 TM. World J Pharm Med Res. 2017;3(9): 343-4.
  4. Vennapusa B, Cruz L, Shah H, Michalski V, Zhang Q. Erythrocyte sedimentation rate measured by the Streck ESR-Auto Plus is higher than with the sediplast westergren method: a validation study. Am J Clin Pathol. 2011;135(3):386-90.
  5. Jou JM, Lewis SM, Briggs C, et al. ICSH review of the measurement of the erythocyte sedimentation rate. Int J Lab Hematol. 2011;33:125‐32.
  6. Westergren A. Studies of the suspension stability of the blood in pulmonary tuberculosis. Acta Med Scand. 1921; 54:247-82.
  7. Patton WN, Meyer PJ, Stuart J. Evaluation of sealed vacuum extraction method (Seditainer) for measurement of erythrocyte sedimentation rate. J Clin Pathol. 1989;42 (3):313-7.
  8. International council for Standardization in Hematology. ICSH recommendation for measurement of ESR. J Clin Pathol 1993;46:198-203.
  9. Patil A, Nayak D, Belurkar SV, Verma S. Validation of an erythrocyte sedimentation rate analyzer with modified westergren method. J Evol Med Den Sci. 2013;2 (3):284-89.
  10. Atas A, Cakmak A, Soran M, Karazeybek H. Comparative study between the Ves‐matic and microerythrocyte sedimentation rate method. J Clin Lab 2008;22:70‐2.
  11. Alexy T, Pais E, Meiselman HJ. A rapid method to estimate westergren sedimentation rates. Rev SciInstrum. 2009; 80(9):096102.
  12. Danesh J, Collins R, Peto R, Lowe GD. Hematocrit, viscosity, erythrocyte sedimentation rate: meta-analyses of prospective studies of coronary heart diseases. Eur Heart J.2000;21:515-20.
  13. AlFadhli S M, Al-Awadhi A M. Comparison of erythrocyte sedimentation rate measurement by the automated SEDI systemTM and conventional westergren method using the Bland and Altman statistical method. Med Princ Pract. 2005;14:241-4.
  14. Hardeman MR, Levitus M, Peliccia A, Bouman A. Test-1 analyzer for determination of ESR. Practical evaluation and comparison with the Westergren technique. Scand J ClinLab 2010; 70:21-5.
  15. Asif N, Uppal R, Mehmood S, Ahmed A, Ali S, Afzal M. Validation of automated ESR methods with conventional method as gold standard. J Islam Med Den Coll2012;(2):81-4.
  16. Hashemi R, Majidi A, Motamed H, Amini A, Najari F, Tabatabaey A. Erythrocyte sedimentation rate measurement using a rapid alternative to the westergren manual method. Emerg. (Tehran). 2015spring; 3(2):50-3.
  17. Sonmez C, Dogan OC, Kaymak AO, Akkaya N, Akin KO, Guntas G. Test-1 analyzer and conventional westergren method for erythrocyte sedimentation rate: A comparative study between two laboratories. J Clin Lab Anal. 2018; 32:e22384.
  18. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet.1986;1:307-10.
  19. Subramanian A, Rangarajan K, Pandey R M, Gandhi J S, Sharma V, and Bhoi S K. Evaluation of an automated erythrocyte sedimentation rate analyzer as compared to the westergren manual method in measurement of ESR. Indian J PatholMicrobiol.2011;54(1):69-73.
  20. Dhruva G, Agravat A, Kakadiya M, Pansuriya H. Automated Erythrocyte Sedimentation Rate Analyzer V/S The westergren manual method in measurement of ESR: a comparative study. Int J Sci Res.2014;3(2):376-8.

Corresponding Author

Dr Anita Harsh

Associate Professor, Department of Pathology, SMS Medical College Jaipur