Title: Correlation of intra ocular pressure, measured by 2 different tonometers

Authors: Prasanthi M, Suganya E, Ravikumar R

 DOI: https://dx.doi.org/10.18535/jmscr/v7i1.100

Abstract

 

Background: Intraocular pressure (IOP) being an modifiable risk factor of glaucoma, has to be assessed earlier for successful identification d and treatment of glaucoma.

Objectives:

  1. To assess the Intra ocular pressure (IOP) using GAT and Schiotz tonometer
  2. To compare and correlate the GAT and Schiotz tonometer in measuring the IOP

Materials and Methods: Cross-sectional study conducted in Ophthalmology department of  SSSIHMS, Andhra Pradesh for a period of 2 months.112 study participants were included in the study and their IOP was assessed using GAT and Schiotz tonometers and the readings were compared. Statistical analysis was done using Pearson correlation and Paired t test.

Results: The mean age of the participants was 53.20 ± 13.03 years. Most of them are males 66(58.9%), illiterates 50(44.7%) and they belonged to Class I socio-economic status 49(43.7%). The average intra ocular pressure was high in females than males measured using both the Goldmann applanation tonometer (GAT) and Schiotz tonometers. The difference in right eye IOP was higher than Left eye. Pearson correlation showed a moderate positive correlation between the 2 tonometric assessment, having r value of 0.577 and it was statistically significant. Paired t test inferred that the difference in the IOP readings assessed by 2 tonometers are not significant.

Conclusion: The insignificant difference between the IOP measured using 2 tonometers, proves that Schiotz tonometer is comparable with gold standard GAT in assessing the IOP, hence Schiotztonometric measurements could be considered as a screening procedure for Glaucoma.

Keywords: Schiotz tonometer,GAT,Glaucoma.

References

  1. American Academy of Ophthalmology, BCSC. Section 10. Glaucoma. 2009-10.
  2. Gordon MA, Beiser JA, Brandt JA, et al. The Ocular Hypertension treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol. 2002; 120(6):714-720.
  3. Frenkel REP, Hong YJ, Shin DH. Comparison of the Tono-Pen to the Goldmann applanation tonometer. Arch Ophthalmol1988,106: 750–53.
  4. Becker-schaffers. Diagnosis and therapy of glaucomas , 8th edition.
  5. Schwartz JT, Dellosso GG. Comparison of Goldmann and Schiotz Tonometry in a Community. Arch Ophthalmol 1966;75 (6):788–795
  6. Schreiber W etal. A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann applanation tonometry, ophtahlmology 2007 Apr;104(4):299-304.
  7. Jonathan SP. Evidence that a gender difference is present from childhood. Ophthalmic and physiological optics. Volume 20, Issue 2, 15 March 2000, Pages 131–136
  8. Vinayak Bhatia, Harsh Kumar. Comparison of the Goldmann-applanation, Schiotz impression &noncontact tonometers in different glaucoma patients & the effect of central corneal thickness. 10th EGS congress, Copenhagen. JUNE 17-22,2012.
  9. Ohana O, Varssano D, Shemesh G. Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty. Indian J Ophthalmol [serial online] 2017 [cited 2018 Dec 5];65:579-83.
  10. Stevens S. How to measure intraocular pressure: Schiötz tonometry. Community Eye Health. 2008;21(66):34.
  11. Armaly MF, Salamoun SG. Schiotz and Applanation Tonometry. Arch Ophthalmol 1963;70(5):603–609.

Corresponding Author

Dr Suganya E

Assistant Professor, Dept. of Community Medicine, TMCH, Chennai, India

Phone: 9894281794; Email: This email address is being protected from spambots. You need JavaScript enabled to view it.